Chattanooga Hylo Table Gas Pelvic Elevation Cylinder
This Chattanooga Hylo Table Gas Pelvic Elevation Cylinder is for the NEWER style Chattanooga ES Hylo Table Gas Pelvic Elevation Cylinder with a starting MFG date of 2016 and Up. It is a direct replacement for the Chattanooga ES Hylo Table Gas Pelvic Elevation Cylinder. This Chattanooga ES Hylo Table Gas Pelvic Elevation Cylinder is a medium install job and will fit Chattanooga ES Hylo Table & Chattanooga ES 2000 Table.
This Includes ONLY Part # 53941 in picture shown
Adjustment for Raising By Itself
1. Extend Breakaway Chest all the way up 2. Loosen Jam Nut 3. Turn Shaft Counter clockwise 1/4 Turn 4. Check for Lock If still slipping turn another 1/4 Round till it doesn't slip, but still locks and unlocks. 5. Tighten Lock Nut back down
Breakaway Adjustment for Not Releasing
1. Loosen Jam Nut
2. Turn Shaft Clockwise 1/4 Turn 3. Check for Lock / Unlock . 4.If not unlocking turn another 1/4 Round till it still locks and unlocks , but doesn't slip. 5. Tighten Lock Nut back down
Warranty Information
This New Chattanooga ES Hylo Table Gas Pelvic Elevation Cylinder is 100% guarantee to be free of defects for one year from date of purchase.